Waltron 9035 sodium analyzer for online sodium monitoring in ultrapure water, with touchscreen interface, automatic calibration and internal datalogger.

help_outlineAsk for Quote

9035 Waltron sodium analyzer

I accept privacy policy rules  

 

9035 sodium analyzer

Waltron 9035 Sodium Analyzer is an online analyzer for sodium measurement in ultrapure-water applications. The official page positions it for common boiler-system measuring points such as make-up water, condensate, boiler feedwater and saturated or main steam.

The instrument includes a color touchscreen interface, true automatic single-point or two-point calibration, grab-sample capability, optional low-flow alarm and quick-connect electrodes. Calibration standards and electrode refill can last up to three months, reducing routine maintenance.

The 9035 covers a wide 0.1 ppb to 10 ppm range, with isolated 4-20 mA outputs, RS-232/RS-485 MODBUS RTU communications, internal datalogger and automatic logging of alarms and calibration results. Its IP65/NEMA 4X stainless-steel enclosure is designed for industrial service.

Reference 9035.
Product Online sodium analyzer for ultrapure water.
Parameter Sodium.
Range 0.1 ppb - 10 ppm.
Accuracy +/-5 % of reading or +/-0.1 ppb, whichever is greater, within +/-5 deg C of calibration temperature.
Reproducibility +/-5 % of reading or +/-0.1 ppb, whichever is greater, at constant temperature.
Response time 90 % of 10-1 ppb step in 3 minutes; 90 % of 100-1 ppb step in 5 minutes.
Temperature compensation Automatic.
pH compensation Alkaline vapor buffer: 0.88 SG ammonia (28 % to 30 % NH3) above 0.1 ppb sodium; 99 % di-isopropylamine below 0.1 ppb.
Digital output RS-232 and RS-485 interfaces using MODBUS RTU protocol.
Analog outputs Two isolated 4-20 mA current outputs.
Alarms Low-flow alarm and three additional voltage-free contacts, 2 A / 250 VAC non-inductive.
Calibration Automatic two-point or single-point; selectable weekly, bi-monthly or monthly frequency.
Datalogger Readings every 10 minutes for 30 days; stores the last 10 alarms and last 10 calibration results.
Sample Temperature 41-131 F / 5-55 deg C; flow 150-400 ml/min; pressure 5-30 psig / 0.35-2.07 bar.
Power 90-250 VAC.
Construction High-strength ABS display, IP65 / NEMA 4X; stainless-steel enclosure, wetted parts stainless steel or PVC.
Dimensions 66.1 x 38.1 x 30.5 cm / 26 x 15 x 12 in.
Weight 15 kg / 33 lb.
Connections Sample inlet 1/4 in OD Swagelok; drain 3/8 in OD tube; eight conduit knockouts.
Distance Up to 100 m on request.
Maintenance Calibration standards and electrode refill can last up to 3 months; quick-connect electrodes.