AP-5000 Advanced Portable Multi-parameter

Aquaread

New

Add to wishlist

The Sensor measures: Water Flow
Cable Length: 3m
Sensor for use in: Water Quality

Standard parameters on the AP-5000

  • Optical DO
  • Specific EC
  • Absolute EC
  • pH
  • ORP
  • TDS
  • Resistivity
  • Salinity
  • SSG
  • Temperature
  • Depth
The AP-5000 is the larger of our advanced portable multiparameter water monitoring probes. It comes with a range of standard sensors included, and much like its smaller counterpart it offers more customisation options that allow you to add extra sensors to the probe. All of our probes are constructed from tough marine grade aluminium that has been hard anodised, making it suitable for use in both fresh and salt waters.

The AP-5000 features 4 Auxiliary sockets to allow you to add further sensors to the probe. Unlike the AP-2000, these auxiliary ports are unrestricted. This means that each port can take either an Ion Selective Electrode (ISE) or an optical sensor. This makes the number of potential sensor combinations vast.

The probe can be used with a variety of cable lengths via the AquaConn Connectors. Cables available in 3, 10, 20 and 30m lengths as standard.

Customisation Options

Customising the AP-5000 is simple, there are 4 ports and you can add any combination of our ISE and optical sensors into these ports.

Probes are supplied with blanking plugs in the auxiliary ports as standard. Simply unscrew these to add your extra electrodes.

Using the AP-5000 as a water level logger

The AP-5000 features a depth sensor. This sensor is capable of measurement accuracy of 0.5%FS. This is very comparable to a dedicated water level logger such as the LeveLine-ECO: 0.2%FS, and the LeveLine (titanium): 0.1%FS.

The depth is zeroed automatically whilst the conductivity is reading zero. As soon as the conductivity registers over 0 the probe depth begins to be calculated. Once the probe is lowered to the desired depth it must be secured, then using the GPS Aquameter the depth can be zeroed once again. Now the AP-5000 will detect small changes in the water level and display them as a positive or negative change.

AP-5000-Spec bigger

Rugged Connection

All Aquaprobes feature the new AquaConn fully metal connectors. The AP-5000 features the all-metal AquaConn connectors. These high specification connectors feature large metallic key for a rigid fit, recessed pins for extra protection during pairing, guide markers on each part of the connector to allow you to instantly push the pair together in the correct alignment.

Rugged AquaConn connector system
 
 
 

Standard Parameters

Oxígeno disuelto Range 0 – 500.0% / 0 – 50.00 mg/L
Resolution 0.1% / 0.01mg/L
Accuracy 0 – 200%: ± 1% of reading. 200% – 500%: ± 10%
Conductividad Range 0 – 200 mS/cm (0 – 200,000 µS/cm)
Resolution 3 Auto-range scales: 0 – 9999 µS/cm, 10.00 – 99.99 mS/cm, 100.0 – 200.0mS/cm
Accuracy ± 1% of reading or ± 1µS/cm if greater
TDS* Range 0 – 100,000 mg/L (ppm)
Resolution 2 Auto-range scales: 0 – 9999mg/L, 10.00 – 100.00g/L
Accuracy ± 1% of reading or ± 1mg/L if greater
Resistencia* Range 5 Ωcm – 1 MΩcm
Resolution 2 Auto-range scales: 5 – 9999 Ωcm, 10.0 – 1000.0 KΩcm
Accuracy ± 1% of reading or ± 1 Ωcm if greater
Salinidad* Range 0 – 70 PSU / 0 – 70.00 ppt (g/Kg)
Resolution 0.01 PSU / 0.01 ppt
Accuracy ± 1% of reading or ± 0.1 unit if greater

Gravedad específica del

agua de mar

Range 0 – 50 st
Resolution 0.1 st
Accuracy ± 1.0 st
pH Range 0 – 14 pH / ± 625mV
Resolution 0.01 pH / ± 0.1mV
Accuracy ± 0.01 pH / ± 5mV
ORP Range ± 2000mV
Resolution 0.1mV
Accuracy ± 5mV
Temperatura Range -5°C – +50°C (23°F – 122°F)
Resolution 0.01° C/F
Accuracy ± 0.5° C

Optional Parameters

Profundidad** Range 0 – 100m (0 – 330F)
Resolution 0.01m/F
Accuracy ± 1%
**Option on the AP-2000. Comes as standard on the AP-7000

Ion Specific Electrodes***

Amónico Range 0 – 9,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L
Accuracy ± 10% of reading
Amoniaco† Range 0 – 9,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 9,000 mg/L
Accuracy ± 10% of reading
Chloride Range 0 – 20,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 20,000 mg/L
Accuracy ± 10% of reading
Flouride Range 0 – 1,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 999.9 mg/L
Accuracy ± 10% of reading
Nitrate Range 0 – 30,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 30,000 mg/L
Accuracy ± 10% of reading
Calcium Range 0 – 2,000mg/L (ppm)
Resolution 2 Auto-range scales: 0.00 – 99.99 mg/L, 100.0 – 2,000 mg/L
Accuracy ± 10% of reading
† Ammonium electrode required. Readings calculated from Ammonium, pH and temperature electrode values.

Optical Electrodes***

Turbidity Range 0 – 3000 NTU
Resolution 2 Auto-range scales: 0.0 99.9 NTU, 100 – 3000 NTU
Repeatability ± 2% of reading
Chlorophyll Range 0 – 500 µg/L (ppb)
Resolution 0.01 µg/L
Repeatability ± 2% of reading
Phycocyanin (Freshwater Blue-Green Algae) Range 0 – 300,000 cells/mL
Resolution 1 cell/mL
Repeatability ± 2% of reading
Phycoerythrin (Marine Blue-Green Algae) Range 0 – 200,000 cells/mL
Resolution 1 cell/mL
Repeatability ± 2% of reading
Rhodamine WT Dye Range 0 – 500 µg/L (ppb)
Resolution 0.01 µg/L
Repeatability ± 5% of reading
Fluorescein Dye Range 0 – 500 µg/L (ppb)
Resolution 0.01 µg/L
Repeatability ± 5% of reading
Refined Oil Range 0 – 10,000 µg/L (ppb)
Resolution 0.1 µg/L
Repeatability ± 2% of reading
CDOM/FDOM Range 0 – 2,000mg/L (ppm)
Resolution 2 Auto-range scales:
0.00 – 99.99 mg/L, 100.0 – 2,000mg/L
Repeatability ± 10% of reading or 2ppm (whichever is greater)

***Note

In addition to the standard parameters supplied with the base Aquaprobe, the AP-2000 has 1 ISE port and 1 optical port, the AP-5000 has 4 ports that can take any combination of ISE or optical sensors and the AP-7000 has 6 ports that can also take any combination of ISE or optical sensors.

The accuracy figures quoted above represent the equipment’s capability at the calibration points at 25°C. These figures do not take into account errors introduced by variations in the accuracy of calibration solutions and errors beyond the control of the manufacturer that may be introduced by environmental conditions in the field.