H6320 Temperature, humidity, CO2 transmitter with two relay and RS232 outputs

H6320

Comet System

New

Temperature, humidity, CO2 level sensor with two output relays.

Included in delivery:

carbon dioxide

  • H6320 sensor
  • Traceable calibration certificate
  • Quick start manual

"

Volume discounts

Quantity Price
(tax excl.)
You Save
%
1+ 465,00 € 0%
5+ 455,70 € 2%
10+ 446,64 € 3.9474%
15+ 437,65 € 5.8816%

465,00 €

tax excl.

Add to wishlist

IP Classification IP30 electronics, IP40 sensors
Communications RS232C
The Sensor measures: Temp. / Rel. Humidity / CO2

The CO2 - carbon dioxide level is recently regarded as an important parameter that substantially determines the quality of the interior climate. Especially in buildings where many people gather, such meeting rooms, hospitals, schools, cinemas, theatres and care centres. With the help of CO2 sensor engineers, technical advisors, environmental experts and health specialists can optimize the ventilation for creation of a healthy interior climate.

Method of CO2 sensor multipoint calibration leads to an excellent accuracy measurements of CO2 in the entire of operating temperature range. With this sensor is the device able to meet the demanding requirements for outdoor use. The measurement principle is based on the NDIR principle  with dual wavelength, which automatically compensates for aging of the sensor. The sensor is resistant against the pollution and provides maintenance-free operation and excellent long-term stability.

Measured temperature and relative humidity values are also converted to other humidity interpretation - dew point temperature, absolute humidity, specific humidity, mixing ratio or specific enthalpy. Degrees Celsius and Fahrenheit are user selectable. 

APPLICATIONS - measuring of temperature, humidity and CO2 level at:

  • building HVAC management
  • climate technology
  • schools, universities
  • meeting rooms, hospitals, cinemas, theatres
  • weather stations

Transmitter is equipped with two relay outputs for alarm indication or control of external devices. Each relay can be assigned to any measured or computed value. For each relay setting of delay, hysteresis, audible alarm is enabled.
Parameters are easy adjustable from regulator keyboard or from the computer.

Transmitter contains a microprocessor based control circuitry in a durable plastic case with connection terminals and temperature an relative humidity sensors in a filter with stainless steel mesh. The CO2 concentration sensor is built inside the case.

Large dual line LCD is an advantage. Display is possible to switch off. Computerized design ensures maximum long term stability and failure indication.

State-of-the-art capacitive polymer sensor ensures excellent calibration long term stability, inertia against water and condensation. Transmitter is designed for use in non-aggressive environment.

Technical parametersValue
Output RS232
Measured Value CO2 + Relative humidity + Temperature
Construction Type Ambient Air
Design Industrial
Temperature Measuring Range -30 to +60°C
Relay Output Yes
Two-State Input No
Lcd Display Yes
PoE No
Maximum switching voltage, current, power of relay output 50V, 2A, 60VA
Audible alarm from built-in beeper - switchable
Range of CO2 concentration measurement 0 to 2000ppm
Accuracy of CO2 concentration measurement ±(50ppm +2% from reading) at 25°C and 1013hPa
Optional range of CO2 concentration measurement 0 to 10000ppm ±(100ppm +5% from reading) at 25°C and 1013hPa
Relative humidity range 0 to 100%
Accuracy of relative humidity measurement ±2.5% relative humidity from 5 to 95% at 23°C
Accuracy of temperature output ±0.4°C
Available temperature units degrees Celsius, Fahrenheit
Accuracy and range of dew point temperature output - for more details see graphs ±1.5°C at ambient temperature T<25°C and RH>30%
range -60 to +80 °C
Accuracy and range of absolute humidity output ±1.5g/m3 at ambient temperature T < 25°C
range 0 to 400 g/m3
Accuracy and range of specific humidity output ±2g/kg at ambient temperature T < 35°C
range 0 to 550 g/kg
Accuracy and range of mixing ratio output ±2g/kg at ambient temperature T < 35°C
range 0 to 995 g/kg
Accuracy and range of specific enthalpy output ±3kJ/kg at ambient temperature T < 25°C
range: 0 to 995 kJ/kg
Measuring interval of CO2 15s
Temperature operating range -30 to +60°C
Filtering ability of sensor cover 0.025mm - filter with stainless steel mesh
Communication protocol ModBus RTU and Advantech ADAM compatible protocol
Communication speed 110 to 115200 Bd
IP protection IP30 electronics, IP40 sensors
Power 9-30Vdc, power consumption approximately 1W
Dimensions 136 x 213 x 45 mm (W x H x D), stem length 75 mm
Weight approximately 360g
Warranty 2 years
Ac/dc adapter 12V/450mA stabilized
code: A1510
Ac/dc adapter 12V/450mA stabilized
plastic sensor protection with stainless filter (gray)
code: F5200
plastic sensor protection with stainless filter (gray)
PTFE sensor protection, white
code: F5300
PTFE sensor protection, white
sintered bronze sensor cover
code: F0000
sintered bronze sensor cover
Tool for easy connection to WAGO terminals 233 for serial output transmitters
code: SP006
Tool for easy connection to WAGO terminals 233 for serial output transmitters
Humidity Standard 10%RH, set 5pcs
code: HM023
Humidity Standard 10%RH, set 5pcs
Humidity Standard 80%RH, set 5pcs
code: HM024
Humidity Standard 80%RH, set 5pcs
Vessel for adjustment and calibration of humidity
code: MD046
Vessel for adjustment and calibration of humidity
Transducer USB/RS232 complete
code: MP006
Transducer USB/RS232 complete

30 other products in the same category: