New
Availability date:
Manual: CTV-B
The Sensor measures: | AC Current |
Output Signal: | 0 to 2,5 VDC |
Cable Length: | 1.8 m |
Power Supply: | Does not require |
Sensor for use in: | Indoor |
Sensors Accessories: | Clamps |
A split-core AC current sensor responsive over the range of 0 to 50 amps AC for use with U12 data loggers with external input channels. With an input current of AC current, sine wave, single phase 50 Hz or 60 Hz, load power factor 0.5 to 1.0 lead or lag.
This sensor is not designed for use with intrinsically-safe (IS) loggers; the molded-plastic housing is rated for indoor use per UL508.
Ranges available (AC Amps): 0-20, 0-50, 0-100, 0-200, 0-600
Accuracy with H8: ±5.0% of full scale
Accuracy with U12: ±4.5% of full scale
Response time (from 10% to 90% of amplitude): see table
Input current: AC current, sine wave, single phase 50 Hz or 60 Hz, load power factor 0.5 to 1.0 lead or lag
Voltage rating: 600 VAC
Construction and operating environment: Molded plastic housing for indoor use per UL508; pivoting leg (leg on CTV-D and CTV-E units is fully removable); UL listed.
Cable Length: 1.8m cable (6 foot)