EC1100 (3200647411)
New
• EC1100 meter
• Electrode stand
• Power adaptor
• Protection cover
Availability date:
With Display | LCD |
Communications | RS232C, USB |
Memory Capacity | 500 |
The Sensor measures: | EC/Res/TDS/Sal/Temp |
- Automatic / manual conductivity calibration up to 4 points
- Adjustable reference temperature and temperature coefficient
- Selectable cell constant and measurement unit
- Automatic / manual temperature compensation
- Temperature calibration function
- Resistivity, total dissolved solids (TDS), and salinity conversions from conductivity
- Four TDS calibration curves for wide range of applications: EN27888, Linear, 442, NaCl
- Two salinity calibration curves: NaCl, Seawater
- Built-in memory can store up to 500 data
- Auto-Hold / Auto-Stable measurement
- RS232C and USB outputs for connecting the meter to a compatible printer or computer
- Meter password setting
- 2 years meter warranty
Model |
LAQUA EC1100 EC/Res/TDS/Sal/Temp (°C) |
---|---|
Condutivity (EC) Range |
0.000 µS/cm to 20.00 mS/cm (k=0.1) 0.00 µS/cm to 200.0 mS/cm (k=1) 0.0 µS/cm to 2.000 S/cm (k=10) |
Resolution |
0.05% of full scale |
Accuracy |
± 0.6% of full scale; ± 1.5% of full scale > 18.0 mS/cm |
Reference Temperature |
15 to 30 °C (Adjustable) |
Temperature Coefficient |
0.00 to 10.00 % (Adjustable) |
Cell Constant |
0.1 / 1.0 / 10.0 (Selectable) |
Measurement Unit |
μS/cm, mS/cm, S/m (Auto-ranging / Manual) |
Calibration Points |
Up to 4 (Auto / Manual) |
TDS Range |
0.00 mg/L to 100 g/L |
Resolution |
0.01 mg/L |
Accuracy |
± 0.1% of full scale |
Calibration Curves |
EN27888 / 442 / Linear (0.40 to 1.00) / NaCl |
Resistivity Range |
0.0 Ω•cm to 2.000 MΩ•cm (k=10) 0.000 kΩ•cm to 20.00 MΩ•cm (k=1) 0.00 kΩ•cm to 200.0 MΩ•cm (k=0.1) |
Resolution |
0.05% of full scale |
Accuracy |
± 0.6% of full scale; ± 1.5% of full scale > 1.80 MΩ•cm |
Salinity Range |
0.00 to 100.00 ppt / 0.000 to 10.000 % |
Resolution |
0.01 ppt / 0.001 % |
Accuracy |
± 0.2% of full scale |
Calibration Curves |
NaCl / Seawater |
Temperature Range |
-30.0 to 130.0 °C |
Resolution |
0.1 °C |
Accuracy |
± 0.4 °C |
Calibration Option |
Yes |
Memory Data |
500 |
Auto-Hold / Auto-Stable |
Yes |
Cell Constant Display |
Yes |
Electrode Status |
On screen display |
Diagnostic Messages |
Yes |
Display |
Custom black and white static LCD |
Inputs |
BNC, phono, DC sockets |
Output |
RS232C, USB |
Power Requirement |
AC adaptor, 100-240V, 50/60Hz |
Weight |
500g |
Dimensions |