pH1300-S (3999960178)
New
Availability date:
With Display | LCD |
Communications | RS232C, USB |
Memory Capacity | 999 |
The Sensor measures: | pH/ORP/Ion/Temp |
Power Supply: | AC adaptor 100 240V, 50/60 Hz |
- LAQUA PH1300 pH/ORP/Ion/Temp custom black and white static LCD meter
- 9615S-10D ToupH glass-body, refillable pH electrode with integrated temperature sensor
- 502-S USA pH buffers: 4.01, 7.00, 10.01 pH buffers & 3.33 M KCl reference electrolyte (250ml each)
- Data acquisition software in USB flash drive (communication cable is sold separately)
- Integrated electrode stand
- Protection cover
- Power adaptor with 6 plugs (US, EU, UK, Australia / New Zealand, Korea, China)
- Instruction manual
Model |
LAQUA PH1300 pH/ORP/Ion/Temperature (°C) |
---|---|
pH Range |
-2.000 to 20.000 pH |
Resolution |
0.1 / 0.01 / 0.001 pH |
Accuracy |
± 0.003 pH |
Calibration Points |
Up to 5 |
Buffer Options |
USA / NIST / Custom |
ORP Range |
± 2000 mV |
Resolution |
0.1 mV |
Accuracy |
± 0.2 mV |
Ion Range |
0.000 µg/L to 9999 g/L |
Resolution |
4 significant digits |
Accuracy |
± 0.3% of full scale |
Calibration Points |
Up to 5 |
Temperature Range |
-30.0 to 130.0 °C |
Resolution |
0.1 °C |
Accuracy |
± 0.4 °C |
Calibration Option |
Yes |
Memory Data |
999 |
Auto Log Data |
Yes |
Real-time Clock |
Yes |
Date / Time Stamp |
Yes |
Auto-Shut Off |
Yes (Programmable: 1 to 30 minutes) |
Auto-Hold / Auto-Stable |
Yes |
Offset / Slope Display |
Yes (Independent acid and alkaline slopes depending on calibration) |
Electrode Status |
On screen display |
Diagnostic Messages |
Yes |
Calibration Alarm |
Yes (Programmable: 1 to 400 days) |
Display |
Custom black and white static LCD |
Inputs |
BNC, phono, DC sockets |
Output |
RS232C, USB |
Power Requirement |
AC adaptor, 100-240V, 50/60Hz |
Weight |
500g |
Dimensions |
170(L) x 174(D) x 73(H) mm |