12962A Portamuestras para Sólidos y Líquidos

12962A, 12963A, 12964A Sample Holders

Solartron

Nuevo

The Model 12962A sample holder is designed to allow accurate impedance tests of solid materials to be performed at room temperature. The sample holder comprises of two parallel electrodes, one of which is fixed in position and the other which can be moved into contact with the sample by adjustment of a micrometer. The sample holder makes use of guard ring and reference techniques in order to improve accuracy by reducing &ldquo:fringing&rdquo: effects at the edge of the sample. The standard electrode provided is 20mm diameter.

Model 12963A electrode kit

This auxiliary electrode kit for the 12962A sample holder allows measurements of different sizes of materials. 10, 30 and40 mmelectrodes are supplied in this kit.

Model 12964A liquid sample holder

The liquid sample holder also makes use of guard ring techniques and is shaped as a container in order to be able to measure small samples of liquid or powders. The sample holder is easily dissembled for cleaning.

The 12962A sample holder is designed to allow accurate impedance tests of solid materials to be performed at room temperature. The sample holder consists of two parallel electrodes, one of which is fixed in position and the other which can be moved into contact with the sample by adjustment of a micrometer. The sample holder makes use of guard ring and reference techniques in order to improve accuracy by reducing “fringing” effects at the edge of the sample. The standard electrode provided is 20mm diameter. 

Model 12963A electrode kit 
This auxiliary electrode kit for the 12962A sample holder allows measurements of different sizes of materials. 10, 30 and 40 mm electrodes are supplied in this kit. 


Model 12964A liquid sample holder 
The liquid sample holder also makes use of guard ring techniques and is shaped as a container in order to be able to measure small samples of liquid or powders. The sample holder is easily dissembled for cleaning. 


Measurement Techniques 

Guard ring 
The 12962A, 12963A and 12964A sample holders make use of a guard ring on the fixed electrode in order to reduce the effect of stray field lines at the edge of the sample which would otherwise lead to measurement errors. 

The guard ring ensures that the electric field lines are parallel throughout the part of the sample which contributes to the impedance measurement. 

Normalized Measurements 
In order to obtain best accuracy from any sample holder, it is advisable to make use of normalization measurement techniques. The sample holder allows normalized measurements to be performed with ease. Typically measurements are taken with the sample material in the cell and then repeated without the sample (using an airgap). The air-gap can easily be set up using the micrometer. The empty cell measurements can then be used to normalize the sample results to further reduce errors due to fringing, connections and instrumentation. 


This sample holder is designed for room temperature measurements only, other sample holders and systems are available for high and low temperature measurements in cryostats and furnaces. 

Electrode diameter 10, 20, 30 or 40 mm
Connections 2-terminal
Sample types solid, liquid and powder
Sample thickness 0.2 mm to 25.4 mm
Sample width and length ≥ 10mm
Impedance range 1 ohm to 100 Tohms

Model 12962A 12963A 12964A Sample Holders Image

Soporte de muestra 12962A:
El portamuestras 12962A está diseñado para permitir la realización de pruebas precisas de impedancia de materiales sólidos a temperatura ambiente. El portamuestras consta de dos electrodos paralelos, uno de los cuales está fijo en su posición y el otro se puede poner en contacto con la muestra mediante el ajuste de un micrómetro. El portamuestras hace uso de anillos de protección y técnicas de referencia para mejorar la precisión al reducir los efectos de "franjas" en el borde de la muestra. El electrodo estándar proporcionado tiene un diámetro de 20 mm.

Kit de electrodos 12963A (requiere 12962A):
Este kit de electrodos auxiliares para el portamuestras 12962A permite mediciones de diferentes tamaños de materiales. En este kit se suministran electrodos de 10, 30 y 40 mm.

Portamuestras de líquido 12964A (requiere 12962A):
El portamuestras de líquido también utiliza técnicas de anillo de protección y tiene la forma de un recipiente para poder medir pequeñas muestras de líquido o polvos. El portamuestras se desmonta fácilmente para su limpieza.

Técnicas de medición

Anillo de guarda:
Los portamuestras 12962A, 12963A y 12964A hacen uso de un anillo de protección en el electrodo fijo para reducir el efecto de líneas de campo perdidas en el borde de la muestra que de otro modo conducirían a errores de medición.
El anillo de protección asegura que las líneas del campo eléctrico sean paralelas en toda la parte de la muestra que contribuye a la medición de impedancia.

Medidas normalizadas:
Para obtener la mejor precisión de cualquier portamuestras, es aconsejable utilizar técnicas de medición de normalización. El portamuestras permite realizar mediciones normalizadas con facilidad. Por lo general, las mediciones se toman con el material de muestra en la celda y luego se repiten sin la muestra (usando un espacio de aire). El espacio de aire se puede configurar fácilmente con el micrómetro. Las mediciones de celda vacía se pueden utilizar para normalizar los resultados de la muestra y reducir aún más los errores debidos a franjas, conexiones e instrumentación.

Este portamuestras está diseñado solo para mediciones de temperatura ambiente; hay otros portamuestras y sistemas disponibles para mediciones de alta y baja temperatura en criostatos y hornos.